Pressure Contribution from Flag Hydrocarbons
40M Lab RGA Scan Results
Job # B22205 Description: Micro D Connectors Date: 2/22/2005
Oven Used: B
AMU 41 4.20E-16 amps from RGA scan listing
AMU 43 6.20E-16 amps from RGA scan listing
AMU 53 4.60E-17 amps from RGA scan listing
AMU 55 6.80E-17 amps from RGA scan listing
AMU 57 4.50E-17 amps from RGA scan listing
Sum Flag H/C AMUs 1.20E-15 amps
Calib leak rate 2.00E-08 torr l/s (Argon)  
AMU 40 (w/leak open) 2.20E-13 amps
AMU 40 (background) 2.20E-14 amps
Calib leak contributes 1.98E-13 amps = (w/leak open) - (background)
Flag H/C Outgassing 1.211E-10 torr l/s = (Sum Flag H/C AMUs) x (Calib leak rate)/(Calib leak contrib.)
Test item surf area cm2
Normalized outgassing #DIV/0! torr l/s-cm2 = Flag H/C Outgassing/Test item surf area
Full description: 10 Male and 10 Female   GlenAir Micro D Connectors
Pre-scan bake: Baked per GlenAir recommended temperature 204C (400F)